HV HIPOT GDC-9560B Sistimi ea matla Insulation Oil Gas Chromatography Analyzer
Ho kenya le ho tlosoa ha kholomo ea chromatographic:
1. Ho kenngoa le ho tlosoa ha kholomo ea chromatographic ho tlameha ho etsoa mocheso oa kamore.
2. Litora tse pakiloeng li na le litiiso tsa ferrule le litiiso tsa gasket.Ho na le mefuta e meraro ea li-ferrules: li-ferrules tsa tšepe, li-ferrules tsa polasetiki le li-graphite ferrules, tseo ho seng bonolo ho li tiisa haholo ha li kenya.Litiiso tsa mofuta oa gasket li hloka li-gaskets tse ncha nako le nako ha kholomo e kenngoa.
3. Hore na lipheletso tse peli tsa kholomo ea chromatographic li kentsoe ka boea ba khalase.Thibela boea ba khalase le ho paka hore li se ke tsa feketsoa ka sebui sa moea ke khase e tsamaisang thepa.
4. Bolelele ba ho kenngoa le ho kenngoa ha kholomo ea capillary ho itšetlehile ka buka ea litaelo ea sesebelisoa.Sebopeho sa likamore tse fapaneng tsa mouoane oa chromatographic se fapane, 'me bolelele ba ho kenya le bona bo fapane.Hoa lokela ho hlokomeloa hore haeba kholumo ea capillary e sebelisoa ka phallo e sa arohaneng, sehokelo pakeng tsa kamore ea mouoane le kholomo e pakiloeng ha ea lokela ho ba le li-probes tse ngata tse hokahaneng le kamore ea mouoane, 'me kholomo ea capillary e lokela ho ba hanyane ho feta cap.
Tšusumetso ea karo-karolelano ea haedrojene ho moea ho detector ea FID:
Ka chromatograph ea khase, karo-karolelano ea hydrogen le moea e lokela ho ba 1:10.Haeba karo-karolelano e phahame haholo, kutloisiso ea "hydrogen flame detector" e tla theoha haholo.Sheba litekanyetso tsa haedrojene le moea oa phallo.Ha khase ea haedrojene le moea e sa lekana, e tla etsa "bang" ho hotetsa, ebe e tima mollo, hangata ha o khantša setofo sa motlakase, oa tima, ebe o hotetsoa hape, ebe o chesa. haedrojene ha ea lekana.
Mokhoa oa ho thibela nale hore e se kobehe:
Batho ba bangata ba qalang ho chromatography hangata ba koba nale le kutu hobane:
1. Sebaka sa ente sa chromatograph ea khase se phunyeletsoa ka thata haholo, 'me se phunyeletsoa ka thata haholo mocheso oa kamore.Ha mocheso oa kamore ea mouoane o ntse o phahama, gasket ea silicone e tla atoloha le ho tiisa.Ka nako ena, ho thata ho phunyeletsa ka sering.
2. Nale e khomaretsoe karolong ea tšepe ea koung ea ente haeba sebaka se se bonolo ho se fumana.
3. Molamu oa syringe o kobehile hobane matla a matla haholo ha o kenya sampuli.Ho na le rack ea sampole bakeng sa chromatographic import.Ha ho khonehe ho koba lere la serinji la setso ha u sebelisa rack ea sampole ho enta lisampole.
4. Kaha lerako le ka hare la sering e silafalitsoe, sutumelletsa lehare la nale hore le khumame nakong ea ente.
5. Chromatograph ea khase e tlameha ho tsitsa nakong ea ente.Haeba u potlakile, u tla koba seringe.Haeba u tloaelane le ts'ebetso ena, e tla phethoa kapele.
Nako ea poso: Jun-07-2022