GDHG-106B CT/PT Analyzer
●Qetella lintho tsohle tsa tlhahlobo ka boeona, ha ho hlokahale ho hokahanya lisebelisoa tse thusang.
●Molaoli ea bohlale o amoheloa.
●Mohatisi o hahelletsoeng ka hare, ho hatisa liphetho habonolo.
●LCD ea skrine e kholo, e bonts'ang ka mokhoa o hlakileng.
●Pontšo ea boiketsetso ea boleng ba ntlha ea lengole, 5% le 10% curve ea phoso.
●Boloka lintlha tsa lihlopha tse 3000 'me ha li lahlehe ka mor'a hore motlakase o theohe.
●Ka koung ea USB.Lintlha li ka bolokoa ka har'a disk ea USB 'me tsa baloa ke PC (faele ea sebopeho sa WORD).
●Moralo o nkehang habobebe, o loketse tlhahlobo ea setšeng.
●Lebokose la polasetiki la boenjiniere le nang le damping, le matla ebile le ka tšeptjoa.
Mesebetsi e ka sehloohong
CT(mofuta oa tšireletso,metering mofuta) | PT |
Vtšobotsi ea khalecurve (ts'ebetso ea thabo) | Vtšobotsi ea khalecurve (ts'ebetso ea thabo) |
Bala ka ho iketsalengole boleng ba ntlha | Bala ka ho iketsalengole boleng ba ntlha |
Ponahatso e iketsang 5% le 10% curve ea liphoso | E fetola tekanyo ea tekanyo |
E fetola tekanyo ea tekanyo | Boikemisetso ba polarity |
Karolelano etekanyo ea bokhopo | Karolelano etekanyo ea bokhopo |
Tekanyo ea mokhahlelo (phapang ea angular) | Tekanyo ea mokhahlelo (phapang ea angular) |
Boikemisetso ba polarity | AC e mamella teko |
Teko ea ente ea mantlha ea hajoale | Ea bobeli methati ya ho hanyetsa |
AC e mamella teko | Ea bobelimoroalo tekanyo |
Ea bobeli methati ya ho hanyetsa | |
Ea bobeli moroalo tekanyo | |
Tlhahlobo e potlakileng (Teko e felletseng ea Iron core demagnetization, thabo e felletseng, tekanyo le tlhahlobo ea polarity ka nako e le 'ngoe) |
Mekhahlelo e ka sehloohong
Ntho | Paramethara | |
Phepelo ea motlakase | AC220V±10%, 50Hz | |
Sephetho | 0-2500Vrms, 5Arms(20Tlhōrō) | |
Sephetho sa hajoale | 0-1000A | |
Nepahalo ya boithatelo | ≤0.5%(0.2% ea palo+ 0.3% ea sebaka) | |
Tekanyo ea bobeli ea ho hanyetsa matsoelintsoeke | Range | 0.1-300Ω |
Ho nepahala | ≤1.0%(0.4% ea palo+ 0.6% ea sebaka) | |
Ea bobeli ea 'netemoroalotekanyo | Range | 5-500 VA |
Ho nepahala | ≤0.5%(0.2% ea palo+ 0.3% ea sebaka)±0.1VA | |
Tekanyo ea phoso ea angular | Ho nepahala | 4min |
Qeto | 0.1 mets | |
CT tekanyo ea karolelano | Range | ≤25000A/5A(5000A/1A) |
Ho nepahala | ≤0.5%(0.2% ea palo+ 0.3% ea sebaka) | |
PTtekanyo ea karolelano | Range | ≤500KV |
Ho nepahala | ≤0.5%(0.2% ea palo+ 0.3% ea sebaka) | |
Tikoloho ea ts'ebetso | Mocheso:-10 ℃-40 ℃,Hbophahamo: ≤90%, Bophahamo: ≤1000m | |
Boima le Boima | Boemo:345mm ×445mm ×380mm Boima ba 'mele:hoo e ka bang 30kg |
Re romelle molaetsa oa hau: